Glove Integrity Tester for Isolators
A solution that adapts to your equipment, from standalone
testing to centralized supervision.
A system tailored to every line
The glove tester enables rapid and reliable detection of perforations using a traceable, reproducible method that complies with current regulations, in particular Annex 1 of GMP. The system can also be adapted to all shapes of glove rings.
Description of the EREA glove tester
The tester developed by EREA is:
Equipped with two sensors that detect the pressure inside the glove
Integrated with a Siemens HMI interface for full control
Designed to be lighter, with lower energy consumption and greater autonomy
Easy to use, with reduced mechanical complexity
Adaptable to different glove port formats (round, oval, etc.)
Featuring exclusive dual-sensor technology
Exclusive dual sensor technology
The EREA glove integrity tester is controlled via the Siemens WinCC Unified system, a recognized reference in the pharmaceutical industry. This interface provides complete and intuitive traceability, including:
Automatic report generation
Alarm log accessible at any time
Secure audit trail to ensure data integrity
This solution is perfectly suited to demanding and regulated environments.
Thanks to its intelligent design, the tester allows continuous calibration using pressure sensors without any disassembly. It incorporates exclusive dual-sensor technology:
The first sensor measures and regulates the pressure inside the glove
The second continuously monitors the reliability of the first sensor
If a deviation is detected, a preventive alarm is triggered to alert the user that calibration is required.
The EREA tester is compatible with WPA2 Enterprise, a centralized authentication protocol ensuring that only authorized devices can connect to the Wi-Fi network. In addition, the Wi-Fi router is configured to be invisible to computers, as it does not broadcast its SSID. This ultra-secure connection is accessible only to users with valid credentials, providing effective protection against cyberattacks.
Operating principle
Thanks to its operating principle based on Hagen-Poiseuille’s law, the tester can simulate leaks equivalent to those generated by very small defects. For example, a calibrated 260 μm orifice used in testing is equivalent, in terms of leak rate, to a hole of around 113 microns in a glove. This means that the tester is sensitive enough to ultimately detect a hole of this size, thus ensuring a high level of micro-leak detection.
Once a year, secure and traceable calibration is carried out using a dedicated programme in the HMI and a test bench that reproduces the test conditions, enabling the accuracy of the pressure sensors to be checked and validated without dismantling, with the option of on-site intervention or return to EREA.
An adapter for every glove port
The adapter is the interface between the tester and the glove port. Its design and the test recipe depend on the geometry and the glove.
Application development
01. Survey the glove port geometry and identify the glove used.
02. Define the matching mechanical interface.
03. Develop the recipe on representative samples.
04. Document the method and its capability under the conditions studied.
Adaptable to all glove port types
Isolators and RABS on the market, subject to a compatibility study.
Round
Oval
ERTP
Custom
EREA offers FAT/SAT, IQ and OQ protocols and services depending on the project. Compatibility is confirmed after assessment.
Recommended application for our Glove Testers
QUALIFICATION & DEPLOYMENT
Design qualification
User requirement specification and functional specification.
Installation qualification
RFID, connectivity, HMI and accessories.
Operational qualification
Method, settings, accuracy, detection threshold, repeatability.
Performance qualification
Routine conditions, operators, glove types and geometries.
FAT/SAT, IQ and OQ protocols and execution depending on project scope.
Standard documentation with GIT One and GIT Link, adaptable with GIT Flex and GIT Fleet.

POSSIBLE APPLICATIONS...
Isolators and RABS in aseptic production
Sterile drug manufacturing
Handling of cytotoxic and highly potent compounds
Routine integrity testing programs
Validation and requalification support
Contamination control strategy (CCS)
TECHNICAL DATA
| FEATURE | VALUE |
| Weight | 2.8 kg (excluding adapter) |
| Dimensions | Tester: 215 × 350 × 121 mm. Overall size with adapter depends on the glove port |
| Pressure sensors | Redundant, ± 0.1% accuracy after calibration |
| Detection threshold | On the order of 100 μm (Hagen-Poiseuille basis), depending on recipe and configuration |
| HMI | Siemens touchscreen |
| Connectivity | Wi-Fi, RFID reader, Active Directory, depending on configuration |
| Power supply | Built-in battery, charging time approx. 1 h 30 min |
| Battery life | Approx. 15 cycles per charge, depending on recipe |
| Cycle Time | Defined by the recipe |
| Calibration | Annual, without disassembly |
| Regulations | EU GMP Annex 1 |
| Traceability | Audit trail, report export, features designed to support 21 CFR Part 11 |
| Data security | User access management, secure HMI |

Recommended accessories for our glove testers
Issues with your Glove Box?
Our trained glove box technician is here to support you. We offer a full range of technical support services, from tailored maintenance plans to complete servicing of your glove box systems.
Have a question or need support? Get in touch with us — we’re happy to help.